Feasibility study of left bundle branch area pacing using simple stylet-driven pacing lead
-
摘要: 目的 探讨无指引鞘条件下单纯使用钢丝支撑的电极导线行左束支区域起搏(LBBAP)的可行性、安全性和起搏特征。方法 总共纳入150例患者,在X线指导下应用单纯钢丝塑形支撑的电极导线行LBBAP。记录术中起搏参数、起搏的QRS波群形态及宽度、左心室达峰时间(LVAT)以及手术并发症等。术后追踪患者进行随访,记录电极参数、心电图和术后并发症情况。结果 150例患者中,LBBAP电极植入成功率为76%,术后QRS宽度(118.43±14.02) ms,术中单极阻抗(589.75±60.14)Ω,双极阻抗(731.88±91.86)Ω,感知(12.25±5.73) mV,阈值(0.85±0.27) V。LVAT为(61.72±10.60) ms。114例患者完成12个月随访,均电极参数稳定,无严重手术相关并发症发生。结论 无指引鞘条件下单纯使用钢丝支撑的电极导线行LBBAP安全可行。Abstract: Objective To explore the feasibility, safety, and pacing characteristics of left bundle branch area pacing(LBBAP) using simple stylet-driven pacing lead without a guide sheath.Methods One hundred and fifty patients were enrolled in this study, and LBBAP was performed with a simple stylet-driven pacing lead without a guide sheath under X-ray guidance. The pacing parameters, paced QRS morphology and width, left ventricular activation time and operation complications were recorded.Results Among 150 patients, the success rate of implantation of LBBAP lead was 76%. The postoperative QRS width was(118.43±14.02) ms, the unipolar impedance was(589.75±60.14) Ω, the bipolar impedance was(731.88±91.86) Ω, the sensing was(12.25±5.73) mV, and the threshold was(0.85±0.27) V, the LVAT was(61.72±10.60) ms. A total of 114 patients completed follow-up for 12 months, with stable electrode parameters and no serious surgical related complications.Conclusion It is safe and feasible to use simple stylet-driven pacing lead LBBAP without a guiding sheath.
-
Key words:
- left bundle branch area pacing /
- stylet-driven pacing lead /
- pacemaker /
- bradycardia
-
表 1 150例患者的基线特征
Table 1. Baseline characteristics of 150 patients
例(%), X ± S 项目 数据 男性 60(40.00) 年龄/岁 67.81±10.69 身高/cm 162.30±7.98 体重/kg 64.56±11.02 病因 SSS 89(59.30) 三度AVB 41(27.30) 高度AVB 9(6.00) 二度二型AVB 9(6.00) 更换电极 2(1.30) 既往史 阵发/持续性心房颤动 13(8.67) 冠心病 41(27.33) 高血压 67(44.67) 糖尿病 25(16.67) 肾功能不全 4(2.67) 术前QRS宽度/ms 110.31±29.33 窄QRS 126(84.00) LBBB 9(6.00) RBBB 15(10.00) 心脏超声指标 LAD/mm 35.80±4.93 LVDD/mm 46.61±3.84 LVEF/% 59.27±5.78 IVST/mm 9.03±1.01 LBBB:左束支传导阻滞;RBBB:右束支传导阻滞;LAD:左心房内径。 -
[1] Chung MK, Patton KK, Lau CP, et al. 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failure[J]. Heart Rhythm, 2023, 20(9): e17-e91. doi: 10.1016/j.hrthm.2023.03.1538
[2] Huang WJ, Su L, Wu SJ, et al. A novel pacing strategy with low and stable output: pacing the left bundle branch immediately beyond the conduction block[J]. Can J Cardiol, 2017, 33(12): 1736. e1-1731736. e3. doi: 10.1016/j.cjca.2017.09.013
[3] Liu X, Gu M, Niu HX, et al. A comparison of the electrophysiological and anatomic characteristics of pacing different branches of the left bundle conduction system[J]. Front Cardiovasc Med, 2021, 8: 781845.
[4] Su L, Wang SJ, Wu SJ, et al. Long-term safety and feasibility of left bundle branch pacing in a large single-center study[J]. Circ Arrhythm Electrophysiol, 2021, 14(2): e009261. doi: 10.1161/CIRCEP.120.009261
[5] De Pooter J, Ozpak E, Calle S, et al. Initial experience of left bundle branch area pacing using stylet-driven pacing leads: a multicenter study[J]. J Cardiovasc Electrophysiol, 2022, 33(7): 1540-1549. doi: 10.1111/jce.15558
[6] Zhang JM, Wang ZF, Zu LN, et al. Simplifying physiological left bundle branch area pacing using a new nine-partition method[J]. Can J Cardiol, 2021, 37(2): 329-338. doi: 10.1016/j.cjca.2020.05.011
[7] Vijayaraman P, Ponnusamy S, Cano Ó, et al. Left bundle branch area pacing for cardiac resynchronization therapy results from the international LBBAP collaborative study group[J]. JACC Clin Electrophysiol, 2021, 7(2): 135-147. doi: 10.1016/j.jacep.2020.08.015
[8] Wang Y, Zhu HJ, Hou XF, et al. Randomized trial of left bundle branch vs biventricular pacing for cardiac resynchronization therapy[J]. J Am Coll Cardiol, 2022, 80(13): 1205-1216. doi: 10.1016/j.jacc.2022.07.019
[9] 王珍, 陈倩, 黄容, 等. 左束支起搏在心力衰竭合并房室传导阻滞患者中的临床疗效初步研究[J]. 临床心血管病杂志, 2022, 38(9): 743-748. doi: 10.13201/j.issn.1001-1439.2022.09.012
[10] Padala SK, Cabrera JA, Ellenbogen KA. Anatomy of the cardiac conduction system[J]. Pacing Clin Electrophysiol, 2021, 44(1): 15-25. doi: 10.1111/pace.14107
[11] Wu SJ, Chen XY, Wang SJ, et al. Evaluation of the criteria to distinguish left bundle branch pacing from left ventricular septal pacing[J]. JACC Clin Electrophysiol, 2021, 7(9): 1166-1177. doi: 10.1016/j.jacep.2021.02.018